GDS-2000A Series User Manual

138

The vertical scale indicator on the
bottom left of the display changes
accordingly for the specific channel.

Range

1mV/div ~ 10V/div (1MΩ). 1-2-5
increments

Stop Mode

In Stop mode, the vertical scale setting can be
changed.

Select Coupling Mode

Panel Operation

1. Press a channel key.

CH1

2. Press Coupling repeatedly to toggle

the coupling mode for the chosen
channel.

Range

DC coupling mode. The whole
portion (AC and DC) of the signal
appears on the display.

AC coupling mode. Only the AC
portion of the signal appears on the
display. This mode is useful for
observing AC waveforms mixed with
DC signals.

Ground coupling mode. The display
shows only the zero voltage level as a
horizontal line.

Example

Observing the AC portion of the waveform using AC
coupling