CONFIGURATION

151

Setup Holdoff Level

Background

The holdoff function defines the waiting period
before the GDS-2000A starts triggering again after
a trigger point. The holdoff function ensures a
stable display if there are a number points in a
periodic waveform that can be triggered. Holdoff
applies to all the triggering types.

Holdoff time

Trigger point

Panel Operation

1. Press the trigger Menu key.

Menu

2. To set the Holdoff time, press the

Holdoff (or Mode/Holdoff) menu
button on the bottom bezel.

3. Use the side menu to set the

Holdoff time.

Range

10ns~10s

Pressing Set to Minimum sets the
Holdoff time to the minimum,
10ns.

Note

Note: The holdoff function is automatically
disabled when the waveform update mode is in
roll mode (page132).