MEASUREMENT

73

Max

Positive peak.

Min

Negative peak.

Amplitude

Difference between the
global high and value and
the global low value,
measured over the entire
waveform or gated region.
(=high − low)

High

Global high voltage. See
page 81 for details.

Low

Global low voltage. See page
81 for details.

Mean

The arithmetic mean value is
calculated for all data
samples as specified by the
Gating option.

Cycle Mean

The arithmetic mean value is
calculated for all data
samples within the first cycle
found in the gated region.

RMS

The root mean square of all
data samples specified by the
Gating option.

Cycle RMS

The root mean square value
is calculated for all data
samples within the first cycle
found in the gated region.

Area

Measures the positive area of
the waveform and subtracts
it from the negative area. The
ground level determines the
division between positive
and negative areas.