Probe specifications, Gtp-070a-4, Gtp-150a-2 – GW Instek GDS-2000A series User Manual User Manual
Page 254
GDS-2000A Series User Manual
254
Probe Specifications
GTP-070A-4
Applicable to: GDS-2072A & GDS-2074A
Position X10
Attenuation Ratio
10:1
Bandwidth
DC to 70MHz
Input Resistance
10MΩ when used with oscilloscopes
with 1MΩ input.
Input Capacitance
28pF~32pF
Compensation Range 25pF~ 45pF
Max. Input Voltage
≤600Vpk, Derating with frequency
Position X1
Attenuation Ratio
1:1
Bandwidth
DC to 6MHz
Input Resistance
1MΩ when used with 1MΩ input
Input Capacitance
120pF~220pF
Max. Input Voltage
≤200Vpk, Derating with frequency
Safety
EN61010-031 CAT II
GTP-150A-2
Applicable to: GDS-2102A & GDS-2104A
Position X10
Attenuation Ratio
10:1
Bandwidth
DC to 150MHz
Rise Time
2.3nS
Input Resistance
10MΩ when used with oscilloscopes
with 1MΩ input.
Input Capacitance
Approximately 17pF
Compensation Range 10 to 35pF
Max. Input Voltage
Position X1
Attenuation Ratio
1:1
Bandwidth
DC to 6MHz
Rise Time
58nS
Input Resistance
1MΩ (Oscilloscope Input)
Input Capacitance
47pF plus oscilloscope capacitance
Max. Input Voltage
300V CAT I, 150V CAT II (DC + peak
AC) derating with frequency.
Safety
EN61010-031 CAT II
500V CAT I, 300Vrms CAT II derating
with frequency.