beautypg.com

B&K Precision 9171-9174 - Manual User Manual

Page 88

background image

76

2. Go down to

EXTERN CONTROL (this option is only

available with DIO/Analog Card installed) and press

or

to select between

OFF, VOLT, or RES. Select VOLT

for external voltage control or

RES for external resistance

control. Select

OFF to disable external analog control.

3. Press to save the change. If

VOLT is selected in the

previous step, go to

EXTERN LEVEL setting right below

EXTERN CONTROL and press

or

to select

between

5V or 10V. Selecting 5V means a 0 – 5V source

will be used to control the full range of the voltage/current.
Similarly,

10V means a 0 – 10V source will be used.

4. Press again to save the change, and press twice

to go to the main screen, which will look like the following
below (Note that voltage and current resolution will be 10 mV
and 10 mA for external analog control):

For Dual Channel models:

OFF

0.00 V 0.00 A

Enter

Enter

Esc