Status – B&K Precision 9171-9174 - Manual User Manual

Page 143

background image

131

STATUS?

Description:

Queries the instrument status

Syntax:

STATUS?

Example:

STATUS?

Return String:

000000

No OVP, OCP, OTP, and other failure or flagged status.

Outputs are off, LCD backlight always ON, and output

mode set to MULTI.

Format:

Refer to the table and description below for details

on the returned data.

When

STATUS?command is used, the power supply will return an

ASCII string of 6 alphanumeric characters, which is a representation
of 3-bytes data encoded in hex. The binary representation of each
byte indicates different statuses of the power supply. See the table
below:

Byte

#

Bits Description

Bit Value Description

2

0-7 System Reserved

-

1

7

6

5

4

3

2

1

0

Channel 1 OVP flag status

Channel 2 OVP flag status

Channel 1 OCP flag status

Channel 2 OCP flag status

Fan failure flag status

AC range error flag status

OTP flag status

NTC flag status

1 is flag, 0 is no flag

1 is flag, 0 is no flag

1 is flag, 0 is no flag

1 is flag, 0 is no flag

1 is flag, 0 is no flag

1 is flag, 0 is no flag

1 is flag, 0 is no flag

1 is flag, 0 is no flag

0

7

6

5

4

3

2

1

0

Channel 1 OVP On/Off status

Channel 2 OVP On/Off status

Channel 1 OCP On/Off status

Channel 2 OCP On/Off status

Channel 1 output On/Off status

Channel 2 output On/Off status

LCD backlight On/Off Status

Output mode Single/Multi status

1 is ON, 0 is OFF

1 is ON, 0 is OFF

1 is ON, 0 is OFF

1 is ON, 0 is OFF

1 is ON, 0 is OFF

1 is ON, 0 is OFF

0 – Always ON, 1 – Timed

0 – MULTI, 1 – SINGLE