B&K Precision 9171-9174 - Manual User Manual
Page 156

144
Description:
Configures/Queries the over voltage protection value of
the selected power supply. See
OVSET/OVSET2
command for similar syntax.
COV2 is for dual channel
models only.
Example:
COV 70.000
Sets the selected power supply’s OVP value to
70.000V.
COV2?
Queries the selected power supply’s second channel
OVP value.
COVP/COVP2
Description:
Configures/Queries the over voltage protection state of
the selected power supply. See
OVP/OVP2 command
for similar syntax.
COVP2 is for dual channel models
only.
Example:
COVP ON
Sets the selected power supply’s OVP state to ON.
COVP2?
Queries the selected power supply’s second channel
OVP state.
COC/COC2
Description:
Configures/Queries the over current protection value of
the selected power supply. See
OISET/OISET2
command for similar syntax.
COC2 is for dual channel
models only.
Example:
COC 3.000
Sets the selected power supply’s OCP value to 3.000 A.
COC2?
Queries the selected power supply’s second channel
OCP value.
COCP/COCP2