IL Set = __.____A on the display. Press to proceed to IML.
If current shunt is used for the calibration, divide the DMM
readings (in DCV) by the resistance (Rs) of the shunt. The result
should be DCV/Rs. Enter this current result into

IL.

4. The power supply will then pre-set another current value for

IML.

Repeat the same process in step 3. Once finished, repeat the
same process for

IMH and IH, as shown below.

5. Once finished, the

OCP Lo calibration will begin. Press

to start. After it’s complete, the unit will select

CALIB OCP Hi for

calibration. Again, after it’s complete, press .

6. Once this calibration process is complete, the calibration data

will be stored into flash memory, and the supply will return to the
calibration menu.

CALIB OCP Hi = Start

IMH Set = X . XXXXA Read = XXXXX

IH Set = X . XXXXA Read = XXXXX

CALIB OCP Lo = Start

Enter

Enter

Enter