B&K Precision 9171-9174 - Manual User Manual
Page 174

162
IL Set = __.____A on the display. Press to proceed to IML.
If current shunt is used for the calibration, divide the DMM
readings (in DCV) by the resistance (Rs) of the shunt. The result
should be DCV/Rs. Enter this current result into
IL.
4. The power supply will then pre-set another current value for
IML.
Repeat the same process in step 3. Once finished, repeat the
same process for
IMH and IH, as shown below.
5. Once finished, the
OCP Lo calibration will begin. Press
to start. After it’s complete, the unit will select
CALIB OCP Hi for
calibration. Again, after it’s complete, press .
6. Once this calibration process is complete, the calibration data
will be stored into flash memory, and the supply will return to the
calibration menu.
CALIB OCP Hi = Start
IMH Set = X . XXXXA Read = XXXXX
IH Set = X . XXXXA Read = XXXXX
CALIB OCP Lo = Start
Enter
Enter
Enter