6 ½ digit dmm – B&K Precision 9171-9174 - Manual User Manual
Page 173

161
Figure 32- Current Calibration with Shunt
Follow the steps below to proceed with current calibration:
1. From the calibration menu, press to enter
CURRENT
CALIBRATION. The following screen will be shown.
2. Select the channel (for dual channel models), then press .
The calibration process will automatically start.
3. The power supply will output a pre-set current to the DMM. Wait
for the DMM reading to settle, then enter it into where it shows
CURRENT CALIBRATION CH: 1
IL Set = X . XXXXA Read = XXX
IML Set = X . XXXXA Read = XXX
Enter
1
6 ½ Digit DMM