161

Figure 32- Current Calibration with Shunt

Follow the steps below to proceed with current calibration:
1. From the calibration menu, press to enter

CURRENT

CALIBRATION. The following screen will be shown.

2. Select the channel (for dual channel models), then press .

The calibration process will automatically start.

3. The power supply will output a pre-set current to the DMM. Wait

for the DMM reading to settle, then enter it into where it shows

CURRENT CALIBRATION CH: 1

IL Set = X . XXXXA Read = XXX

IML Set = X . XXXXA Read = XXX

Enter

1

6 ½ Digit DMM