B&K Precision 9171-9174 - Manual User Manual
Page 178

166
Figure 35- Dual Channel External Analog Input Calibration
2. Connect the 10VDC source to the analog inputs on the
DIO/Analog Interface Card as shown in Figure 36 for single
channel.
Figure 36- Single Channel Analog Control Calibration Setup
3. Set the DC source to output 1.000 V and wait for approximately
5 seconds until the output voltage is stable. Press to finish
the first step.
4. Next, change the output of the DC source to 9.000 V and follow
the same as step 3.
CH 1 EXT . CURR Res = Start
CH 2 EXT . VOLT Res = Start
CH 2 EXT . CURR Res = Start
Calib EXT . Lo Volt = 1 . 000 V
Calib EXT . Lo Volt = 9 . 000 V
CH 1 EXT . VOLT Res = Start
Enter