beautypg.com

B&K Precision 9171-9174 - Manual User Manual

Page 138

background image

126

Format:


CURRent/CURR2

Note:

CURR2 command is for dual channel models only.

Description:

Configures/Queries the ISET current setting. For dual

channel models, use CURR2 command for channel 2.

Syntax:

CURRent

CURR2

CURRent?

CURR2?

- 0 – max. rated current output

Example:

CURR2 1.234

Sets the second channel ISET current to 1.234 A on a

dual channel model.

Query:

CURR?

Return String:

1.234

Format:


ERRor?

Description:

Queries system errors

Syntax:

ERRor?

Example:

ERR?

Return String:

0

Refer to the table below for error description.

Format:

Error Code Description
0

No Error

1

Command Error

2

Execution Error

3

Query Error

4

Input Range Error




This manual is related to the following products: