External analog input calibration – B&K Precision 9171-9174 - Manual User Manual
Page 177

165
External Analog Input Calibration
This calibration option is only available if the optional
DIO/Analog
Interface Card is installed in the power supply.
Follow the steps below to proceed with external analog input
calibration.
1. From the calibration menu, press to enter
EXT. INPUT
CALIB. If the single channel DIO/Analog Interface Card is
installed, then Figure 34 will be shown. If the dual channel
version is installed, Figure 35 will be displayed.
Figure 34- Single Channel External Analog Input Calibration
CH 1 EXT . CURR Res = Start
Calib EXT . Lo Volt = 1 . 000 V
Calib EXT . Lo Volt = 9 . 000 V
CH 1 EXT . VOLT Res = Start
3