61


WARNING:

DO NOT output 10A or more with only one pair of

(+) and (-) terminals in the rear panel. Both (+)

terminals and both (-) terminals must all be

connected for applications requiring more than

10A output. Each terminal can only accept

maximum of 10A. Exceeding this may damage the

power supply.


+S + + - - - S

Rear Panel Output

Remove Shorting bars

+S + + - - - S

DUT

+ -

+S + - -S

Remove shorting bars

Front Panel Output

Rear Panel Output

Figure 19- Remote Sense Setup for 10A or Higher Current Output Operation