beautypg.com

B&K Precision 9171-9174 - Manual User Manual

Page 163

background image

151

drop, use remote sense terminals to compensate and output a
more accurate voltage to the DUT.

Q: On my dual channel power supply, how do I set a timer
output for each channel?

- The two channels are dependent on the same timer function.

Therefore, when used it will affect both outputs.

Q: I want to use the external analog control from the
DIO/Analog interface card. However, I only want to adjust
voltage.

- The external analog control requires that both voltage and

current analog controls are connected to either an external DC
voltage or resistance. If only the voltage input terminals are
connected, current will remain set to 0 A, and thus output will be
0. Two sources (DC voltage or resistance) must be used
simultaneously to function properly; one for voltage and one for
current.

Q: I cannot set voltage or current at the maximum rating.

- Go into the menu and check the

OUTPUT SETTINGS to verify

that the

VOLT LIMIT SETTING and CURR LIMIT SETTING are

configured within the range you want to adjust.

- Verify the protection settings under the

PROTECTION menu to

make sure OVP and OCP settings are configured to the
maximum value or greater than the voltage and current value
you want to adjust.


This manual is related to the following products: