Configures/Queries the over current protection state of
the selected power supply. See
OCP/OCP2 command
for similar syntax.
COCP2 is for dual channel models
only.
Example:
COCP ON
Sets the selected power supply’s OCP state to ON.
COCP2?
Queries the selected power supply’s second channel
OCP state.
CMODE?/CMODE2?
Description:
Queries the output mode (CV or CC) of the selected
power supply.
CMODE2? Is for dual channel models
only.
Query:
CMODE?
Queries channel 1 for dual channel models, or single
channel.
CMODE2?
Queries channel 2 for dual channel models.
Return String:
CV
Format:
CV or CC
Examples
Here are a few examples of how to configure/query an instrument
connected in the RS-485 chain.
-
Query the IDN information for power supply with address 3
CADR 3
CIDN?
-
Configure ISET current of a dual channel supply with
address 10 to 1.234 A.
CADR10
CPC2 1.234
Return String: OK