PROG 1

PROG:RUN ON

Select program 1

Use RUN ON command to execute

Example 2:

To create an output sequence like above, send the following list of
commands in order:

PROG 2

Choose program number

PROG:CLE

Clear program 2 data

PROG:REP 0

No repeat after running this program

PROG:TOTA 8

Set program 2 to have 8 steps in total

PROG:STEP 1

Settings for step 1

PROG:STEP:CURR 2

Set output current to 2 amperes

PROG:STEP:VOLT 20

Set output voltage to 20 volts

PROG:STEP:ONT 0.5

Set output ON time to 0.5 sec

PROG:STEP 2

Choose step 2

PROG:STEP:CURR 2

PROG:STEP:VOLT 15

PROG:STEP:ONT 0.5

PROG:STEP 3

Settings for step 3

PROG:STEP:CURR 2

PROG:STEP:VOLT 20

PROG:STEP:ONT 0.5

PROG:STEP 4

Choose step 4

PROG:STEP:CURR 2

PROG:STEP:VOLT 10

PROG:STEP:ONT 0.5

PROG:STEP 5

Choose step 5