Voltage Calibration

- Use the setup diagram shown in Figure 33 below for voltage

calibration. Connect the rear output terminals of the power

supply to the DMM.

Figure 33- Voltage Calibration Setup


Follow the steps below to proceed with voltage calibration:
1. From the calibration menu, press to enter

VOLTAGE

CALIBRATION. The following screen will be shown.

2. Select the channel (for dual channel models), then press .

The calibration process will automatically start.

VOLTAGE CALIBRATION CH: 1

VL Set = X . XXXXV Read = XXXX

VML Set = XX . XXXXV Read = XXXXX

Enter

2

6 ½ Digit DMM