167

5. To begin calibration with the external resistor, remove the DC

source and connect the precision resistor to the control card, as
shown in Figure 37(right) below.

Figure 37- External Resistance Control Calibration setup (left) Current (right)

Voltage

6. The supply will display

CH1 EXT. VOLT Res. Press to

calibrate.

7. Once finished,

CH1 EXT. CURR Res will be selected. Connect

the precision resistor as shown in Figure 37(left). Press to
proceed with the calibration.

8. (Dual channel) The calibration process for the dual channel

DIO/Analog Interface Card is similar to the above steps for
single channel, with the exception of repeating steps 3-7 twice.

9. (Dual channel) For the external voltage calibration, connect the

leads like Figure 38 shown below.

Enter

Enter