-81-

*1) Forward-direction signals are taken in with forward total time measurement, while reverse-direction signals are taken in with reverse

total time measurement. They are conducted alternately for the transmission count. Forward and reverse signal data is added for the
transmission count and averaged. The result is 1 output of signal in forward/reverse direction.

Window
(WID)

Ultrasonic signal
(SIG)

Transmission start
(TRG)

A
transmission signal

B
receieved signal

Measurement window

Window open

Trigger

Part B magnified
Received signal

Part A magnified
Transmission signal

[Outline drawing of signal processing]

Transmission
voltage

Transmission pattern

(Example: Burst 3)

Trigger level

0V

0.75V

1.0V

-0.75V
-1.0V

Detection point

Forward total time measurement

Reverse total time measurement

Signal peak

Upstream side
Sensor
transmission

Downstream side
Sensor reception

Upstream side
Sensor reception

Downstream side
Sensor
transmission

Transmission interval

Note) Make the setting, following the description in “6.6.6. Checking received waveforms”.

Operation
(example)

Set measurement method to “METHOD 1”.

Key operation

Description

Display

ź

Press the

key 4 times to display “MAINTENANCE MODE”.

MAINTENANCE MODE

ENT

ź

Press the

ENT

key once to display “RAS INFORMATION”.

RAS INFORMATION
0000000000000000

ź

Press the

key 14 times to display “DETAILS”.

DETAILS

ENT

ź

Press the

ENT

key once to display “TRANS.COUNT”.

TRANS.COUNT

128

ź

Press the

key 4 times to display “MEAS.METHOD”.

MEAS.METHOD

METHOD:2

ENT

ź

Press the

ENT

key once to select, and press it once again to blink on the

2nd line.

MEAS.METHOD

METHOD:2

ź

Press the

key twice to display “METHOD 1”.

MEAS.METHOD

METHOD:1

Press the

ENT

key once to register.

MEAS.METHOD

COMPLETE

Ļ

ENT

ź
ź
ź
ź

ʊʊʊ “METHOD 1” has been registered. ʊʊʊ

MEAS.METHOD

METHOD:1

ESC

Press the

ESC

key twice and then press the

key once to enter the

measurement mode.

0.000 m/s
0.000 m3/h

Bulletin F-107-UXF3