Checking sending/receiving, Bulletin f-107-uxf3 – Dwyer UXF3 User Manual

Page 128

background image

-120-

6.6.6.2. Checking sending/receiving

Monitor the waveform, and check the status of received waveform.

a) Normal status

Transmitted/received waveform for

both forward and reverse directions

CH1 : 500mV/div
CH2 : 5V/div

Point
1. Check that overall noise level is kept at 0.3

Vp-p or lower.

Relationship between the received

waveform and the sampling window

CH1 (SMP)

CH2 (SIG)

Magnified

Noise level

Pipe inside echoic wave (generated only
in case of V method mounting).
In mounting by the Z method, there is no
echoic wave.

Forward direction:

Sends from the upstream of the
detector, and receives at the
downstream.

Reverse direction:

Reverse procedure of forward direction.

Sampling window

CH2 (WID)

CH1 (SIG)

CH1 : 500mV/div
CH2 : 5V/div

Point
1. There is the received waveform

for

SMP of

CH1

in front of the center of LOW.

2. The received waveform is approximately

0.75Vo-p.

Magnified view of signals

CH1 : 500mV/div

Point
1. Startup is kept within 3 to 6 waves.
2. The peak (amplitude) does not fluctuate.

If the peak fluctuates vertically, air may be
mixed in.

3. The time base must not fluctuate.

If it does, there may be influence by turbulent
flow or drift current.

The received waveform controls the peak to be approximately 1.5Vp-p.

Peak fluctuation

Time base fluctuation

Forward direction

Forward direction

Reverse direction

Received waveform

Transmitted waveform

Bulletin F-107-UXF3