4.10. Application operation of parameter

4.10.1. Setting automatic 2 ranges

Description
z The function carries out a measurement while changing over the range

according to the flow rate.

z The current output changes with the action range as illustrated on the right.
z The hysteresis can be set to between 0 and 20% of the smaller range.
z Upon setting DO1, DO2 or DO3 to "FULL SCALE 2", a contact outputs "FULL

SCALE 2" action. Select “ACTIVE ON” or “ACTIVE OFF” separately. (See
4.10.5.)

z Settable range: 0.3 to 32 [m/s]; 0.984 to 105 [ft/s] in terms of flow velocity for

FULL SCALE1 and FULL SCALE2.
* Preset PIPE PARAMETER and FLOW UNIT.
* If a value beyond the settable range is inputted, "INPUT ERROR" appears

and then last setting is resumed.

* If "FLOW UNIT" has been changed after setting the range, redo the range

setting.

* When FULL SCALE2 is not used (in the case of single range), set “0” to

FULL SCALE2.

The flow rate unit is as selected by "FLOW UNIT". Before setting range, set the "FLOW UNIT". (See 4.8.2.)

For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand. (See Section 4.4.1.)

Operation
(example)

Set “AUTO 2” to “RANGE TYPE”, 10[m

3

/h] to “FULL SCALE1”, and 60[m

3

/h] to “FULL SCALE2”.

Set “RANGE HYS.” to 7%.
* Preset “PIPE PARAMETER” and “FLOW UNIT”.

Key operation

Description

Display

ź

Press the

key twice to display “OUTPUT SETTING”.

OUTPUT SETTING

ENT

ź

Press the

ENT

key once to display “ZERO ADJUSTMENT”.

ZERO ADJUSTMENT

SET ZERO

ź

Press the

key 4 times to display “RANGE”.

RANGE

ENT

ź

Press the

ENT

key twice to blink the cursor.

RANGE TYPE

SINGLE

ź

Press the

key once, and select “AUTO 2”.

RANGE TYPE

AUTO 2

ENT

ź

Press the

ENT

key once to display “RANGE TYPE”.

RANGE TYPE

AUTO 2

ź

Press the

key once to display “FULL SCALE1”.

FULL SCALE1

20.0000 m3/h

ENT

ź

Press the

ENT

key once to blink the cursor on the 2nd line.

FULL SCALE1

0

020.0000 m3/h

ź

Press the

key several times to align the cursor to “2”.

FULL SCALE1

00

2

0.0000 m3/h

ź

Press the

key several times to change to “1”.

Note) To change the decimal point position, align the cursor with a place

to change to, and press the

key.

FULL SCALE1

00

1

0.0000 m3/h

Press the

ENT

key once to register.

FULL SCALE1

COMPLETE

Ļ

ENT

ź
ź
ź
ź

ʊʊʊ FULL SCALE1 has been registered. ʊʊʊ

FULL SCALE1

10.0000 m3/h

ź

Press the

key once to display “FULL SCALE2”.

FULL SCALE2

0.0000 m3/h

20mA

4mA

FULL SCALE1

FULL SCALE2

Hysteresis

Bulletin F-107-UXF3