Application operation of parameter, Setting automatic 2 ranges, Bulletin f-107-uxf3 – Dwyer UXF3 User Manual
Page 57
![background image](/manuals/371886/57/background.png)
-49-
4.10. Application operation of parameter
4.10.1. Setting automatic 2 ranges
Description
z The function carries out a measurement while changing over the range
according to the flow rate.
z The current output changes with the action range as illustrated on the right.
z The hysteresis can be set to between 0 and 20% of the smaller range.
z Upon setting DO1, DO2 or DO3 to "FULL SCALE 2", a contact outputs "FULL
SCALE 2" action. Select “ACTIVE ON” or “ACTIVE OFF” separately. (See
4.10.5.)
z Settable range: 0.3 to 32 [m/s]; 0.984 to 105 [ft/s] in terms of flow velocity for
FULL SCALE1 and FULL SCALE2.
* Preset PIPE PARAMETER and FLOW UNIT.
* If a value beyond the settable range is inputted, "INPUT ERROR" appears
and then last setting is resumed.
* If "FLOW UNIT" has been changed after setting the range, redo the range
setting.
* When FULL SCALE2 is not used (in the case of single range), set “0” to
FULL SCALE2.
For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand. (See Section 4.4.1.)
Operation
(example)
Set “AUTO 2” to “RANGE TYPE”, 10[m
3
/h] to “FULL SCALE1”, and 60[m
3
/h] to “FULL SCALE2”.
Set “RANGE HYS.” to 7%.
* Preset “PIPE PARAMETER” and “FLOW UNIT”.
Key operation
Description
Display
ź
Press the
key twice to display “OUTPUT SETTING”.
OUTPUT SETTING
ENT
ź
Press the
ENT
key once to display “ZERO ADJUSTMENT”.
ZERO ADJUSTMENT
SET ZERO
ź
Press the
key 4 times to display “RANGE”.
RANGE
ENT
ź
Press the
ENT
key twice to blink the cursor.
RANGE TYPE
SINGLE
ź
Press the
key once, and select “AUTO 2”.
RANGE TYPE
AUTO 2
ENT
ź
Press the
ENT
key once to display “RANGE TYPE”.
RANGE TYPE
AUTO 2
ź
Press the
key once to display “FULL SCALE1”.
FULL SCALE1
20.0000 m3/h
ENT
ź
Press the
ENT
key once to blink the cursor on the 2nd line.
FULL SCALE1
0
020.0000 m3/h
ź
Press the
key several times to align the cursor to “2”.
FULL SCALE1
00
2
0.0000 m3/h
ź
Press the
key several times to change to “1”.
Note) To change the decimal point position, align the cursor with a place
to change to, and press the
key.
FULL SCALE1
00
1
0.0000 m3/h
Press the
ENT
key once to register.
FULL SCALE1
COMPLETE
Ļ
ENT
ź
ź
ź
ź
ʊʊʊ FULL SCALE1 has been registered. ʊʊʊ
FULL SCALE1
10.0000 m3/h
ź
Press the
key once to display “FULL SCALE2”.
FULL SCALE2
0.0000 m3/h
20mA
4mA
FULL SCALE1
FULL SCALE2
Hysteresis
Bulletin F-107-UXF3