Installation, Installation location of flow transmitter, Caution – Dwyer UXF3 User Manual
Page 16
-8-
3. INSTALLATION
Select an installation location that satisfies the following conditions for ease of maintenance and inspection, service life of
the instrument, and assurance of reliability all considered.
(1) A location where ambient temperature and humidity are -20 to +55°C and 90% RH or less for transmitter
(UXF3), -20 to +80°C and 90% RH or less for detector (SX1) and -20 to +60°C and 90% RH or less for
detector (SX2).
(2) A location not exposed to direct sunlight nor inclement weather.
(3) Space for periodic inspection and wiring work is available.
(4) A location not subjected to radiated heat from a heating furnace, etc.
(5) A location not subjected to corrosive atmosphere.
(6) A location not to be submerged.
(7) A location free from excessive vibration, dust, dirt and moisture.
3.1. Installation location of flow transmitter
Secure at least 100 mm (3.94 in.) of space between the flow transmitter and nearby wall. Also secure a space of opening the front
cover in case of maintenance.
Allow space for cable wiring under the case.
OP
EN
147(5.79)
>100
(3.94)
147(5.79)
400(15.75) or more
Fig. 3.1 Top view of mounting (Flow transmitter : UXF3) [unit: mm(inch)]
CAUTION
Bulletin F-107-UXF3