4.10.7. How to compensate the measurement value

Description
z Measurement value can be calibrated arbitrarily.

Zero point and span adjustment can be made.
Settable range
(1) Zero point : -5 to +5 [m/s]; -16.4 to +16.4 [ft/s] in terms

of flow velocity through the piping.

(2) Span

: ±200%

The output value (reading, analog output and total output)
is computed by the following expression.

Output =

Measurement value × [Span set value %]

100

+ Zero point

For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand. (See Section 4.4.1.)

Operation
(example)

Compensate the zero point to 0.5m

3

/h, and the span by +1%.

Key operation

Description

Display

ź

Press the

key twice to display “OUTPUT SETTING”.

OUTPUT SETTING

ENT

ź

Press the

ENT

key once to display “ZERO ADJUSTMENT”.

ZERO ADJUSTMENT

SET ZERO

ź

Press the

key 10 times to display “CALIBRATION ZERO”.

CALIBRATION ZERO

0.000 m3/h

ENT

ź

Press the

ENT

key once to blink the cursor.

CALIBRATION ZERO

0

0000.000 m3/h

ź

Press the

key 6 times to move the cursor.

CALIBRATION ZERO

00000.

0

00 m3/h

ź

Press the

key 5 times to set “5”.

CALIBRATION ZERO

00000.

5

00 m3/h

Press the

ENT

key once to register.

CALIBRATION ZERO

COMPLETE

Ļ

ENT

ź
ź
ź
ź

ʊʊʊ “CALIBRATION ZERO” has been registered. ʊʊʊ

CALIBRATION ZERO

0.500 m3/h

ź

Press the

key once to display “CALIBRATION SPAN”.

CALIBRATION SPAN

100.0 %

ENT

ź

Press the

ENT

key once to blink the cursor.

CALIBRATION SPAN

1

00.0 %

ź

Press the

key twice to move the cursor.

CALIBRATION SPAN

10

0

.0 %

ź

Press the

key once to set “1”.

CALIBRATION SPAN

10

1

.0 %

Press the

ENT

key once to register.

CALIBRATION SPAN

COMPLETE

Ļ

ENT

ź
ź
ź
ź

ʊʊʊ “CALIBRATION SPAN” has been registered. ʊʊʊ

CALIBRATION SPAN

101.0 %

ESC

Press the

ESC

key once and then press the

key 3 times to enter

the measurement mode.

0.000 %
0.000 m3/h

Output

0

Zero adjustment movement

Output

0

Span movement

100%

Flow rate

Flow rate