-79-

4.11.10. Initializing setting parameters

Description
z Initializes the setting parameters saved in the memory.
z Initializes those other than the zero adjusted values or analog output calibration value.

Initialize code: 0100 (4-digit number)

z This parameter is intended for our service personnel.
z Do not attempt to initialize the setting parameters. Otherwise measurement is disabled.

When the parameter is initialized, display language is set to English.
To switch the display language, refer to “4.5. Display language”.

For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand. (See Section 4.4.1.)

Operation
(example)

Initializes the setting parameters.

Key operation

Description

Display

ź

Press the

key 4 times to display “MAINTENANCE MODE”.

MAINTENANCE MODE

ENT

ź

Press the

ENT

key once to display “RAS INFORMATION”.

RAS INFORMATION
0000000000000000

ź

Press the

key 11 times to display “MEMORY INITIAL”.

MEMORY INITIAL

ENT

ź

Press the

ENT

key twice to blink on the 2nd line.

MEMORY INITIAL

0

000

ź

Set “0100” by the

and the

key.

MEMORY INITIAL

0

1

00

Press the

ENT

key once to register.

MEMORY INITIAL

COMPLETE

Ļ

ENT

ź
ź
ź

ʊ Flow transmitter is reset, and the measurement mode is displayed. ʊ

0.000 m/s
0.000 m3/h

CAUTION

Bulletin F-107-UXF3