6.6.2. Displaying the data in maintenance mode

Follow the procedure shown below to check possible display errors.

Key operation

Description

Display

ź

Press the

key 4 times to display “MAINTENANCE MODE”.

MAINTENANCE MODE

ENT

ź

Press the

ENT

key once to display “RAS INFORMATION”.

RAS INFORMATION
0000000000000000

ź

Press the

key 13 times to display “DATA DISPLAY”.

DATA DISPLAY

ENT

ź

Press the

ENT

key.

z Displays the transit time and the window value calculated from

the piping setting.

T0 C: 89 usec
WinC: 80 usec

ź

Press the

key once.

z Displays the measurement value of transit time, T1 (forward

time), and T2 (reverse time) from the piping setting.

T1: 0.000 usec
T2: 0.000 usec

ź

Press the

key once.

z Displays the measurement value of average transit time, T0,

and transit time difference, DT.

T0: 0.000 usec
DT:PUGE

ź

Press the

key once.

z Displays the calculated value of pass time of the substances

other than fluid, Ta, and angle of incidence of the fluid, T.

Ta: 0.0000 usec
Tf: 0.000°

ź

Press the

key once.

z Displays the calculation value of sound velocity in fluid, Cf, and

Reynolds number, Re.

Cf: 0.0 m/s
Re: 0

ź

Press the

key once.

z Displays correction coefficient of flow velocity distribution, K,

and flow velocity, V.

K: 1.3333
V: 0.000 m/s

ź

Press the

key once.

z Displays the intensity of received signals.

The larger the value, the larger the intensity of received
signals.
Normal measurement values fall in 35% or more.
If the display appears as 0%, no signals are being received.
Ultrasonic waves may not be transmitted because of
insufficient water volume or rust of piping.

AGC U: 0.00 %
AGC D: 0.00 %

ź

Press the

key once.

z Displays the peak value of received signal waveform.

Normal values stably fall within the range from 5528 to 6758.
If the value fluctuates significantly, objects that constitute
barriers against ultrasonic wave transmission such as air
bubbles or foreign matter may be contained in the fluid.
Stop the flow and check if normal value is resumed.
If so, there is a possibility that air bubbles are contained.

P/H U: 6143
P/H D: 6143

ź

Press the

key once.

z Displays the detection level value of received signal waveform.

TRG U: 25.00%
TRG D: 25.00%

ESC

Press the

ESC

key or the

key to display the measurement

mode.