Bulletin f-107-uxf3 – Dwyer UXF3 User Manual
Page 51
![background image](/manuals/371886/51/background.png)
-43-
4.9.2.4. Determining how to dispose of total at error (BURNOUT)
Description
BURNOUT (TOTAL)
z Determines how to dispose of the total when the measurement status is abnormal on account of an empty pipe interior or bubbles
mixed in fluid (common to total indication and total pulse output).
z Settable range:
HOLD
: Stops the total (as factory set).
NOT USED : Continues the total according to a flow rate marked immediately before the error occurrence.
BURNOUT TIMER
z Sets the time from error occurrence to error processing.
z Settable range: 0 to 900sec (factory set: 10sec)
The total continues until the burnout timer is actuated.
For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand. (See Section 4.4.1.)
Operation
(example)
Change the processing from "BURNOUT" to "HOLD", and change the burnout timer setting from 10 seconds to 15
seconds.
Key operation
Description
Display
ź
Press the
key twice to display “OUTPUT SETTING”.
OUTPUT SETTING
ENT
ź
Press the
ENT
key once to display “ZERO ADJUSTMENT”.
ZERO ADJUSTMENT
SET ZERO
ź
Press the
key 5 times to display “TOTAL”.
TOTAL
ENT
ź
Press the
ENT
key once to display “TOTAL MODE”.
TOTAL MODE
START
Press the
key 4 times to display “BURNOUT(TOTAL)”.
Because HOLD (factory set) is already registered, go to the next step.
BURNOUT(TOTAL)
HOLD
ź
ź
ź
ź
Note) For setting “NOT USED”, press the
key, and the
key to
select “NOT USED”.
ź
Press the
key once to display “BURNOUT TIMER”.
BURNOUT TIMER
10sec
ENT
ź
Press the
ENT
key once to blink the cursor.
BURNOUT TIMER
0
10sec
ź
Press the
key twice to move the cursor.
BURNOUT TIMER
01
0
sec
ź
Press the
key 5 times to set “5”.
BURNOUT TIMER
01
5
sec
Press the
ENT
key once to register.
BURNOUT TIMER
COMPLETE
Ļ
ENT
ź
ź
ź
ź
ʊʊʊ BURNOUT TIMER has been registered. ʊʊʊ
BURNOUT TIMER
15sec
ESC
Press the
ESC
key twice and then press the
key 3 times to enter
the measurement mode.
0.00 m3/h
+ 0.00 m3
ENT
Bulletin F-107-UXF3