Zero adjustment, Bulletin f-107-uxf3 – Dwyer UXF3 User Manual

Page 40

background image

-32-

4.7. Zero Adjustment

Description
z Zero point is calibrated.

Settable range:

CLEAR

: Clears the zero point calibration value to "0".

Used in case the flow cannot be stopped when calibrating the zero point.

Note 1) Where possible, stop the flow and carry out "SET ZERO" stated below.

Otherwise, an error may occur in the zero point.

SET ZERO : A point where "SET ZERO" is set will be regarded as zero flow. The flow should be stopped with a full pipe condition

when calibrating the zero point.

Note 2) The flow must completely be stopped.

Otherwise, the actual flow rate value upon completing "set zero" becomes a constant reading offset error .

It takes ten seconds to several tens of seconds to complete adjustment, depending on pipe diameter.

For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand. (See Section 4.4.1.)

Operation
(example)

Completely fill the piping, close the upstream and downstream valves, and proceed to zero point calibration.

Key operation

Description

Display

ź

Press the

key twice to display “OUTPUT SETTING”.

OUTPUT SETTING

ENT

ź

Press the

ENT

key twice to display “ZERO ADJUSTMENT” and blink the

cursor.

ZERO ADJUSTMENT

CLEAR

ź

Press the

key once, and select “SET ZERO”.

ZERO ADJUSTMENT

SET ZERO

Press the

ENT

key once to carry out “SET ZERO”.

* Be sure to completely stop the flow beforehand.

ZERO ADJUSTMENT

COMPLETE

Ļ

ENT

ź
ź
ź
ź

ʊʊʊ Zero adjustment has been completed. ʊʊʊ

ZERO ADJUSTMENT

SET ZERO

ESC

Press the

ESC

key once, and the

key 3 times to enter the

measurement mode.

0.000 m/s
0.000 m3/h