4.9. Output Setting

4.9.1. Setting of flow rate range

4.9.1.1. Setting of flow rate range (single range)

Description
z The range (full scale) of flow rate to be measured is set.

* The analog output (4-20mA) corresponds to the range setting.

z Settable range: 0.3 to 32 [m/s]; 0.984 to 105 [ft/s] in terms of flow velocity in piping

* The piping parameters and FLOW UNIT must be set beforehand.
* If a value beyond the settable range is inputted, "INPUT ERROR" appears and then last

setting is resumed.

* If "piping parameters" or "FLOW UNIT" has been changed after setting the range,

recommence the range setting.

The flow rate unit is as selected by "FLOW UNIT" in the "MEASURE SETUP" mode.

(Refer to Section 4.8.2.)

For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand. (See Section 4.4.1.)

Operation
(example)

Set 60m3/h to range type, SINGLE/FULL SCALE1.
* Set the piping parameters and "FLOW UNIT" beforehand.

Key operation

Description

Display

ź

Press the

key twice to display “OUTPUT SETTING”.

OUTPUT SETTING

ENT

ź

Press the

ENT

key to enter the “ZERO ADJUSTMENT” mode.

ZERO ADJUSTMENT

SET ZERO

ź

Press the

key 4 times to display “RANGE”.

RANGE

ENT

ź

Press the

ENT

key once to display “RANGE TYPE”.

Because SINGLE (factory set) is already registered, go to the next step.

RANGE TYPE

SINGLE

ź

Press the

key once to display “FULL SCALE1”.

FULL SCALE1

15.000 m3/h

ENT

ź

Press the

ENT

key once to blink the cursor.

FULL SCALE1

0

0015.000 m3/h

ź
ź
ź
ź

Move the cursor by the

key, and change the numeric value by the

key.

000

1

5.000 m3/h

000

6

5.000 m3/h

0006

5

.000 m3/h

ź
ź
ź
ź

Change the full scale to “60”.
Note) To change the decimal point position, align the cursor with a place

to change to and press the

key likewise.

FULL SCALE1

000006

0

.0 m3/h

Press the

ENT

key once to register.

FULL SCALE1

COMPLETE

Ļ

ENT

ź
ź
ź
ź

ʊʊʊ FULL SCALE1 has been registered. ʊʊʊ

FULL SCALE1

60.000 m3/h

ESC

Press the

ESC

key 3 times and then press the

key 3 times to

enter the measurement mode.

0.000 m/s
0.000 m3/h

20mA

4mA

0

100%

Flow rate

Full scale 1

Bulletin F-107-UXF3