Installation of flow transmitter, Wall mounting transmitter, Pipe mounting transmitter – Dwyer UXF3 User Manual
Page 20: Wall mounting (flow transmitter : uxf3), 2b pipe stand mounting (flow transmitter : uxf3)
![background image](/manuals/371886/20/background.png)
-
12-
3.3. Installation of flow transmitter
The flow transmitter may be mounted on a wall or 2B pipe stand (option).
3.3.1. Wall mounting (Flow transmitter : UXF3)
For wall mounting, use two M8 bolts.
Drill holes according to the mounting hole dimensions shown below, and fasten the flow transmitter using the M8 bolts.
3.3.2. 2B pipe stand mounting (Flow transmitter : UXF3)
When mounting on 2B pipe, be sure to use a complete set of fixtures (U bolt, support fixture, plain washer,
spring washer, nut) furnished if optionally designated. Tighten the nut by hand. If any support fixture is not used
or if the assembly is excessively tightened by tool, the wall mounting fixture may be damaged.
Mount the instrument on 2B pipe stand as illustrated below.
Nominal Standard tightening torque
M8
12.5 [Nm]
2B pipe stand
U bolt (M8)
Plain washer
Spring washer
Nut
Support fixture
Bulletin F-107-UXF3
[unit: mm]