4.10.5.2. How to validate the alarm output
Description
z Select a contact output as DO1 and/or DO2 when received wave or E
2
PROM is abnormal.
z Settable range
ALL
: Select a contact output when hardware and received wave (nothing, unstable) are abnormal.
HARDWARE FAULT : Select a contact output when circuit is abnormal.
PROCESS ERROR : Select a contact output when received wave is abnormal.
For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand. (See Section 4.4.1.)
Operation
(example)
Set the DO1 output to “PROCESS ERROR”.
Also, set the contact to “ACTIVE ON”.
Key operation
Description
Display
ź
Press the
key twice to display “OUTPUT SETTING”.
OUTPUT SETTING
ENT
ź
Press the
ENT
key once to display “ZERO ADJUSTMENT”.
ZERO ADJUSTMENT
SET ZERO
ź
Press the
key 6 times to display “DO1 OUT”.
* Press the
key again to display “DO2 OUT”.
* Press the
key once again to display “DO3 OUT”.
DO1 OUT
NOT USED
ENT
ź
Press the
ENT
key once to blink the cursor.
DO1 OUT
NOT USED
ź
Press the
key 4 times to display “ALARM” on the 2nd line.
DO1 OUT
ALARM
ENT
ź
Press the
ENT
key once to display the ALARM select panel.
ALARM
ALL
ź
Press the
key twice to display “PROCESS ERROR”.
ALARM
PROCESS ERROR
Press the
ENT
key once to register.
ALARM
COMPLETE
Ļ
ENT
ź
ź
ź
ź
ʊʊʊ “PROCESS ERROR” has been registered. ʊʊʊ
STATUS OUT
CONTACT ACTION
ENT
ź
Press the
ENT
key once to display “CONTACT ACTION”.
CONTACT ACTION
ACTIVE ON
Press the
ENT
key once to register “ACTIVE ON”(normally off).
* To select normally on, press the
key.
CONTACT ACTION
COMPLETE
Ļ
ENT
ź
ź
ź
ź
ʊʊʊ “ACTIVE ON” has been registered. ʊʊʊ
STATUS OUT
CONTACT ACTION
ESC
Press the
ESC
key twice and then press the
key 3 times to enter
the measurement mode.
0.000 m/s
0.000 m3/h