Bulletin f-107-uxf3 – Dwyer UXF3 User Manual
Page 73
-65-
4.10.6.2. How to validate the total preset with the external contact.
Description
z The total value becomes the preset value by closing or opening the contact.
z The contact should be closed or open for about 1 second.
z When total presetting, “TOTAL PRESET” is indicated on the 2nd line of the LCD display (for about 4 seconds).
z Related setting items: 4.9.2.2. Setting the preset value”, “4.9.2.3. TOTAL mode”
Note 1) This function is valid when the LCD display is measurement screen. When the display is setting screen, it becomes invalid.
For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand. (See Section 4.4.1.)
Operation
(example)
Set the DO1 output to “TOTAL RESET”.
Key operation
Description
Display
ź
Press the
key twice to display “OUTPUT SETTING”.
OUTPUT SETTING
ENT
ź
Press the
ENT
key once to display “ZERO ADJUSTMENT”.
ZERO ADJUSTMENT
SET ZERO
ź
Press the
key 9 times to display “NOT USED” on the 2nd line.
DI1 INPUT
NOT USED
ENT
ź
Press the
ENT
key once to blink the cursor.
DI1 INPUT
NOT USED
ź
Press the
key once to display “TOTAL RESET” on the 2nd line.
DI1 INPUT
TOTAL RESET
Press the
ENT
key once to register “TOTAL RESET”.
DI1 INPUT
COMPLETE
Ļ
ENT
ź
ź
ź
ź
ʊʊʊ “TOTAL RESET” has been registered. ʊʊʊ
DI1 INPUT
TOTAL RESET
ESC
Press the
ESC
key once and then press the
key 3 times to enter
the measurement mode.
0.000 %
0.000 m3/h
Bulletin F-107-UXF3