4.10.8. Setting of the operation mode
Description
z Used to switch computation cycle and output cycle.
z Settable range
NORMAL
: Standard mode (factory-set value), computation/output cycle is approximately 0.5 seconds.
HIGH SPEED : High speed response mode, computation/output cycle is approximately 0.2 seconds.
For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand. (See Section 4.4.1.)
Operation
(example)
Switch the operation mode to the high speed response mode.
Key operation
Description
Display
ź
Press the
key twice to display “OUTPUT SETTING”.
OUTPUT SETTING
ENT
ź
Press the
ENT
key once to display “ZERO ADJUSTMENT”.
ZERO ADJUSTMENT
SET ZERO
ź
Press the
key 12 times to display “OPERATION MODE”.
OPERATION MODE
NORMAL
ENT
ź
Press the
ENT
key once to blink the cursor.
OPERATION MODE
NORMAL
ź
Press the
key 6 times to move the cursor.
OPERATION MODE
HIGH SPEED
Press the
ENT
key once to register.
OPERATION MODE
COMPLETE
Ļ
ENT
ź
ź
ź
ź
ʊʊʊ “OPERATION MODE” has been registered. ʊʊʊ
OPERATION MODE
HIGH SPEED
ESC
Press the
ESC
key once and then press the
key 3 times to enter
the measurement mode.
0.000 %
0.000 m3/h
Reference
The difference between standard mode and high speed mode:
High speed mode is unfit for the measurement when foreign objects or air bubbles are contained.
Standard mode is about 10 times more resistant to entry of foreign objects or air bubbles than high speed mode.