background image

-47-

4.9.5. Setting the damping

Description
z Used for attenuating the variation of measured value.

A time constant is set (response time of about 63%).

Settable range: 0.0 to 100.0sec in 0.1 sec steps

Note) In case you set to 0 sec, response time becomes as below.

x System cycle 0.2sec
x Dead time 0.2sec or less, time constant 0.1sec

For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand. (See Section 4.4.1.)

Operation
(example)

Change the damping from 5 to 20 sec.

Key operation

Description

Display

ź

Press the

key twice to display “OUTPUT SETTING”.

OUTPUT SETTING

ENT

ź

Press the

ENT

key once to display “ZERO ADJUSTMENT”.

ZERO ADJUSTMENT

SET ZERO

ź

Press the

key once to display “DAMPING”.

DAMPING

5.0 sec

ENT

ź

Press the

ENT

key once to blink the cursor.

DAMPING

0

05.0 sec

0

0

5.0 sec

0

2

5.0 sec

02

5

.0 sec

ź

Set “20” by the

key and the

key.

DAMPING

02

0

.0 sec

Press the

ENT

key once to register.

DAMPING

COMPLETE

Ļ

ENT

ź
ź
ź
ź

ʊʊʊ DAMPING has been registered. ʊʊʊ

DAMPING

20 sec

ESC

Press the

ESC

key once and then press the

key 3 times to enter

the measurement mode.

0.000 m/s
0.000 m3/h