4.10.6. Setting the DI input
Description
z Zero adjustment or total preset can be performed by no-voltage contact input signal.
Note 1) To use the DI input, communication board (option) is required.
z Settable range
NOT USED
: Contact input is not used.
TOTAL RESET
: Total value becomes the preset value.
ZERO ADJUSTMENT : Zero adjustment can be performed.
CONTACT ACTION
ACTIVE ON
: Normally off. Activated when a contact is closed.
ACTIVE OFF
: Normally on. Activated when a contact is open.
For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand. (See Section 4.4.1.)
4.10.6.1. Invalidating the DI input
Description
z Select not to use the contact input of the DI1 INPUT.
For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand. (See Section 4.4.1.)
Operation
(example)
Change the DI1 setting from “ZERO ADJUSTMENT” to “NOT USED”.
Key operation
Description
Display
ź
Press the
key twice to display “OUTPUT SETTING”.
OUTPUT SETTING
ENT
ź
Press the
ENT
key once to display “ZERO ADJUSTMENT”.
ZERO ADJUSTMENT
SET ZERO
ź
Press the
key 9 times to display “NOT USED” on the 2nd line.
DI1 INPUT
ZERO ADJUSTMENT
ENT
ź
Press the
ENT
key once to blink the cursor.
DI1 INPUT
ZERO ADJUSTMENT
ź
Press the
key once to display “NOT USED” on the 2nd line.
DI1 INPUT
NOT USED
Press the
ENT
key once to register “NOT USED”.
DI1 INPUT
COMPLETE
Ļ
ENT
ź
ź
ź
ź
ʊʊʊ “NOT USED” has been registered. ʊʊʊ
DI1 INPUT
NOT USED
ESC
Press the
ESC
key once and then press the
key 3 times to enter
the measurement mode.
0.000 %
0.000 m3/h