4.11. MAINTENANCE MODE

4.11.1. How to calibrate the analog output

Description
z The calibration is performed so as to obtain 4mA and 20mA when the analog signal (4-20mA DC) output is 0% and 100%,

respectively.

z Connect an ammeter to Iout terminals as shown below. In the CURRENT CALIBRATION mode, select 4mA or 20mA, and operate the

key (UP) or the

key (Down).

+

-

Ammeter

12

13

GND HF2

11

HF1

10

GND

9

8

-

+

For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand. (See Section 4.4.1.)

Operation
(example)

Adjust the 4mA and 20mA analog outputs.

Key operation

Description

Display

ź

Press the

key 4 times to display “MAINTENANCE MODE”.

MAINTENANCE MODE

ENT

ź

Press the

ENT

key once to display “RAS INFORMATION”.

RAS INFORMATION
0000000000000000

ź

Press the

key once to display “CURRENT”.

CURRENT

CALIBRATION

ENT

ź

Press the

ENT

key twice to enter the calibration mode of 4mA output.

Adjust the output to 4mA by the

(UP) and the

(down) key, while

observing the output of calibration devices such as an ammeter.

CALIBRATION

4 mA

Press the

ENT

key once to register the adjustment result.

CALIBRATION

COMPLETE

Ļ

ENT

ź
ź
ź
ź

ʊʊʊ 4mA adjustment result has been registered. ʊʊʊ

CALIBRATION

4 mA

ź

Press the

key once, and select 20mA.

CALIBRATION

20mA

ENT

ź

Press the

ENT

key twice to enter the calibration mode of 20mA output.

Adjust the output to 20mA by the

(UP) and the

(down) key.

CALIBRATION

20mA

Press the

ENT

key once to register the adjustment result.

CALIBRATION

COMPLETE

Ļ

ENT

ź
ź
ź
ź

ʊʊʊ 20mA adjustment result has been completed. ʊʊʊ

CALIBRATION

20mA

ESC

Press the

ESC

key twice and then press the

key once to enter the

measurement mode.

0.000 %
0.000 m3/h

Bulletin F-107-UXF3