Avionics Software
4.2
Frequency Transients
4.2.1
Maximum Ramp Rate
100Hz/s up, 200Hz/s down
Test procedure TBD
4.2.2
Transients Outside of Steady State Range
Per Figure 6
Test procedure TBD
4.3
DC Content
Per Figure 11
Test procedure TBD
SUPP
Supplementary Transient Test Requirements
Tests 1-7, Table 3.3.2-3A
Tests performed at 360, 530, and 800 Hz
LSUPP-1
SUPP
Supplementary Transient Test Requirements
Tests 8-14, Table 3.3.2-3A
Tests performed at 360, 530, and 800 Hz
LSUPP-2
SUPP
Supplementary Transient Test Requirements
Tests 15-20, Table 3.3.2-3A
Tests performed at 360, 530, and 800 Hz
LSUPP-3
SUPP
Supplementary Transient Test Requirements
Tests 21-27, Table 3.3.2-3A
Tests performed at 360, 530, and 800 Hz
LSUPP-4
SUPP
Supplementary Transient Test Requirements
Tests 28-34, Table 3.3.2-3A
Tests performed at 360, 530, and 800 Hz
LSUPP-5
SUPP
Supplementary Transient Test Requirements
Tests 35-40, Table 3.3.2-3A
Tests performed at 360, 530, and 800 Hz
LSUPP-6
SUPP
Supplementary Transient Test Requirements
Tests 41-47, Table 3.3.2-3A
Tests performed at 360, 530, and 800 Hz
LSUPP-7
SUPP
Supplementary Transient Test Requirements
Tests 48-53, Table 3.3.2-3A
Tests performed at 360, 530, and 800 Hz
LSUPP-8
SUPP
Supplementary Transient Test Requirements
Tests 54-62, Table 3.3.2-3A
Tests performed at 360, 530, and 800 Hz
LSUPP-9
SUPP
Supplementary Transient Test Requirements
Tests 63-74, Table 3.3.2-3A
Tests performed at 360, 530, and 800 Hz
LSUPP-10
SUPP
Supplementary Transient Test Requirements
Tests 1-6 Table 3.3.2-3B
Tests performed at 360, 530, and 800 Hz
LSUPP-11
SUPP
Supplementary Transient Test Requirements
Tests 7-12 Table 3.3.2-3B
Tests performed at 360, 530, and 800 Hz
LSUPP-12
SUPP
Supplementary Transient Test Requirements
Tests 13-15 Table 3.3.2-3B
Tests performed at 360, 530, and 800 Hz
LSUPP-13
4.4.3.2
230VAC:
Test No.
Section
Subject
Test Limits
Comment
File Ref.
1.0
Normal AC Steady State
1.1
Individual Phase Voltage
190.0 to 244.0 Volts RMS
30 mins, 800Hz and 360Hz
HVAC-11
1.2
Average of Three-Phase Voltages
193.0 to 241.0 Volts RMS
30 mins, 800Hz and 360Hz
HVAC-12
1.3
Phase Displacement
120.0 +/- 4.0°
Apply from main gui screen manually
1.4
Phase Voltage Unbalance
Max. 12.0 Volts RMS
Apply each test 30 mins
HVAC-14
1.5
AC Voltage Modulation
1.5.1
Maximum Amplitude
10.0 Volts Peak-to-Valley
Follows specification Figure 1
HVAC-15
1.5.2
Frequency Components
Per Figure 1
Figure only, not a test. (Refer to spec)
1.6
Voltage Waveform
1.6.1
Total Harmonic Content
Max. 10% of Fundamental
30 mins each test
HVAC-16-1
1.6.2
Individual Harmonic Content
Max. 8% of Fundamental
Requires audio generator
1.6.3
DC Content
+/-0.20 Volts (after first cycle)
30 mins each test
HVAC-16-2
1.6.4
Crest Factor
1.41 +/- 0.15