Table a(cf)230 – AMETEK Compact i/iX Series Software Manual User Manual
Page 42
User Manual
– Rev H
California Instruments
42
Avionics Software
3.4.1.7
Table A(CF)230:
Test No.
Section
Subject
Test Limits
Comment
File Ref.
16.5.1
Normal Operating Conditions (ac)
16.5.1.1b(1) 1 Phs Voltage and Frequency (ac)
200V to 244V, 390Hz to 410Hz
Apply for 30 mins each test
EACF230_VFNORM1P
16.5.1.1c(1) 3 Phs Voltage and Frequency (ac)
200V to 244V, 390Hz to 410Hz
Apply for 30 mins each test
EACF230_VFNORM3P
16.5.1
Emergency Operating Conditions (ac)
16.5.1.1b(2) 1 Phs Voltage and Frequency (ac)
200V to 244V, 360Hz to 440Hz
Apply for 30 mins each test
EACF230_VFEMER1P
16.5.1.1c(2) 3 Phs Voltage and Frequency (ac)
200V to 244V, 360Hz to 440Hz
Apply for 30 mins each test
EACF230_VFEMER3P
16.5.1
Normal Operating Conditions (ac)
16.5.1.2 Voltage Modulation (ac)
Vamp = 0.36V to 2.48V, Mrate = 1Hz to 200Hz
Figure 16-1
EACF230_VMOD
16.5.1.3 Frequency Modulation (ac)
Famp = 4.9 Hz to 0.2Hz, Mrate = 0.014Hz to 100Hz Figure 16-2
EACF230_FMOD
16.5.1.4b Momentary Power Interruptions (ac)
Table 16-1
Equipment with Digital Circuits
EACF230_PINTB
16.5.1.4c Momentary Power Interruptions (ac)
Other Equipment
EACF230_PINTC
16.5.1.5 Normal Transients (ac)
16.5.1.5.1 - Normal Voltage Surge (ac)
320V to 140V, 400 Hz
EACF230_NVSUR
16.5.1.5.2 - Normal Frequency Surge (ac) 230V, 350Hz to 400Hz
EACF230_NFSUR
16.5.1.7 Voltage DC Content (ac)
16.5.1.7 - 1 Phs Voltage DC Content (ac)
± 0.2VDC
Apply for 30 mins each test
EACF230_DCOFS1P
16.5.1.7 - 3 Phs Voltage DC Content (ac)
± 0.2VDC
Apply for 30 mins each test
EACF230_DCOFS3P
16.5.1.8 Voltage Distortion (ac)
8% +2%/-0% VTHD - Clipped Sine, 400Hz
Apply for 30 mins each test
EACF230_VTHD
16.5.2
Abnormal Operating Conditions (ac)
16.5.2.1b 1 Phs Voltage (ac)
194V to 268V, 400Hz
Apply for 5 mins each test
EACF230_VABN1P
16.5.2.1c 3 Phs Voltage (ac)
194V to 268V, 400Hz
Apply for 5 mins each test
EACF230_VABN3P
16.5.2.1d 1 Phs Voltage and Frequency (ac)
200V to 244V, 370Hz to 430Hz
Apply for 5 mins each test
EACF230_VFABN1P
16.5.2.1e 3 Phs Voltage and Frequency (ac)
200V to 244V, 370Hz to 430Hz
Apply for 5 mins each test
EACF230_VFABN3P
16.5.2.2 Momentary Undervoltage (ac)
20V and 120V, 400Hz
Apply for 7 seconds each test
EACF230_VUND
16.5.2.3 Abnormal Transients (ac)
16.5.2.3.1 - Abnormal Voltage surge
296V to 360V, 400Hz
Figure 16.5 - apply at 10 sec intervals EACF230_AVSUR
16.5.2.3.1 - Abnormal Frequency transients 0V to 230V, 320Hz to 480Hz
Figure 16.5 - apply at 10 sec intervals EACF230_AFSUR