Lx/ls series –a350 option limitations – AMETEK Compact i/iX Series Software Manual User Manual

Page 136

background image

User Manual

– Rev H

California Instruments

136

Avionics Software

6.1.3

Lx/Ls Series

–A350 Option Limitations

The extent of coverage of the

–A350 option as implemented on the Lx/Ls Series platform is shown in

the table below. Tests marked with an „X‟ generally require additional hardware to be used. Tests
marked with a „P‟ require arbitrary waveform capability, which is available on Lx models or Ls models
with the

–ADV option only.

Note that the Lx/Ls Series only provides AC output so none of the DC tests (LDC Power Group and
DC Offset tests in AC power groups) are supported on the Lx/Ls Series. External DC supplies per
Figure B.16 and Figure B.17 are required to perform these tests.

Section 6.2.1.4 and Appendix D current harmonics measurements will require the use of an external
power analyzer per Figure B.14 of ABD0100.1.8.1 since the iX Series measurement system does not
support the required measurement bandwidth for this test.

Note also that f

or EUT‟s with a nominal voltage of 230Vrms, the optional 200/400 output voltage

range, option

–EHV is required to perform some of the voltage transient tests that require an output

voltage higher than 300Vrms AC. (Test 102 and 202 for 230V power groups) The available current on
the

–EHV range is lower compared to the standard 300V (Lx) or 270V (Ls) range so the Lx/Ls power

level should be sized accordingly to support the EUT‟s to be tested.