beautypg.com

Table a(wf)115 – AMETEK Compact i/iX Series Software Manual User Manual

Page 37

background image

User Manual - Rev H

California Instruments

Avionics Software

37

3.4.1.3

Table A(WF)115:

Test No.

Section

Subject

Test Limits

Comment

File Ref.

16.5.1

Normal Operating Conditions (ac)

16.5.1.1b(1) 1 Phs Voltage and Frequency (ac)

100V to 122V, 360Hz to 800Hz

Apply for 30 mins each test

EAWF115_VFNORM1P

16.5.1.1c.2(1) 3 Phs Voltage and Frequency (ac)

100V to 122V, 360Hz to 800Hz

Apply for 30 mins each test

EAWF115_VFNORM3P

16.5.1

Emergency Operating Conditions (ac)

16.5.1.1c.2(2) 3 Phs Voltage and Frequency (ac)

100V to 122V, 360Hz to 800Hz

Apply for 30 mins each test

EAWF115_VFEMER3P

16.5.1

Normal Operating Conditions (ac)

16.5.1.2

Voltage Modulation 360Hz (ac)

Vamp = 0.18V to 1.24V, Mrate = 1Hz to 200Hz

Figure 16-1

EAWF115_VMOD360

16.5.1.2

Voltage Modulation 800Hz (ac)

Vamp = 0.18V to 1.24V, Mrate = 1Hz to 200Hz

Figure 16-1

EAWF115_VMOD800

16.5.1.3

Frequency Modulation 360Hz (ac)

Famp = 4.9 Hz to 0.2Hz, Mrate = 0.014Hz to
100Hz

Figure 16-2

EAWF115_FMOD360

16.5.1.3

Frequency Modulation 800Hz (ac)

Famp = 4.9 Hz to 0.2Hz, Mrate = 0.014Hz to
100Hz

Figure 16-2

EAWF115_FMOD800

16.5.1.4b

Momentary Power Interruptions (ac)

Table 16-1 360Hz nom

Equipment with Digital Circuits

EAWF115_PINTB360

16.5.1.4b

Momentary Power Interruptions (ac)

Table 16-1 800Hz nom

Equipment with Digital Circuits

EAWF115_PINTB800

16.5.1.4c

Momentary Power Interruptions (ac)

360Hz nom

Other Equipment

EAWF115_PINTC360

16.5.1.4c

Momentary Power Interruptions (ac)

800Hz nom

Other Equipment

EAWF115_PINTC800

16.5.1.4d

Momentary Power Interruptions (ac)

Table 16-2

Additional Requirement (Var Freq)

EAWF115_PINTD

16.5.1.5

Normal Transients (ac)

16.5.1.5.1 - Normal Voltage Surge (ac) 160V to 70V, 360 Hz

EAWF115_NVSUR360

16.5.1.5.1 - Normal Voltage Surge (ac) 160V to 70V, 800 Hz

EAWF115_NVSUR800

16.5.1.6

Normal Frequency Surge (ac)

115V, 360Hz to 800Hz

EAWF115_NFSUR

16.5.1.7

Voltage DC Content (ac)

16.5.1.7 - 1 Phs Voltage DC Content
(ac)

± 0.1VDC (360Hz nom)

Apply for 30 mins each test

EAWF115_DCOFS1P360

16.5.1.7 - 1 Phs Voltage DC Content
(ac)

± 0.1VDC (800Hz nom)

Apply for 30 mins each test

EAWF115_DCOFS1P800

16.5.1.7 - 3 Phs Voltage DC Content
(ac)

± 0.1VDC (360Hz nom)

Apply for 30 mins each test

EAWF115_DCOFS3P360

16.5.1.7 - 3 Phs Voltage DC Content
(ac)

± 0.1VDC (800Hz nom)

Apply for 30 mins each test

EAWF115_DCOFS3P800

16.5.1.8

Voltage Distortion (ac)

10% +2%/-0% VTHD - Clipped Sine, 360Hz

Apply for 30 mins each test

EAWF115_VTHD360

16.5.1.8

Voltage Distortion (ac)

10% +2%/-0% VTHD - Clipped Sine, 800Hz

Apply for 30 mins each test

EAWF115_VTHD800

16.5.2

Abnormal Operating Conditions (ac)

16.5.2.1b

1 Phs Voltage (ac)

97V to 134V, 360Hz

Apply for 5 mins each test

EAWF115_VABN1P360

16.5.2.1b

1 Phs Voltage (ac)

97V to 134V, 800Hz

Apply for 5 mins each test

EAWF115_VABN1P800

16.5.2.1c

3 Phs Voltage (ac)

97V to 134V, 360Hz

Apply for 5 mins each test

EAWF115_VABN3P360