beautypg.com

AMETEK Compact i/iX Series Software Manual User Manual

Page 15

background image

User Manual - Rev H

California Instruments

Avionics Software

15

Control / Display

Description

Revert to
Steady State
on Abort

When selected, the output of the power source will be set back
to the steady state settings in effect before the start of the test
when the operator aborts a test. Normally, the output of the
source at the end of a test run is determined by the settings in
the test sequence file loaded.

Options

This section contains additional options that can be set by the operator as
needed. The following options are available:

Auto Acquire
EUT Data
after
execution

When set, this option causes the GUI program to initiate an EUT
performance data measurement sequence automatically at the
end of a test (normal mode) or test step (single step mode).

Show Phase
B, C data

For three-phase EUTs, data for phases B and C can be
displayed as well. In most cases, the settings for all three
phases are the same and there is no need to display them all as
it takes up screen space. However, this is not the case for
voltage imbalance tests. This is a display feature only and has
no effect on the actual test. Note however that if the power
source is in single-phase mode, settings for phase B and C will
be ignored. The phase mode of the power source can only be
changed from the main GUI screen.

External
Trigger
Output

An external trigger signal is provided through the trigger
out connection during test execution. There are two modes
of operation that can be selected:
BOT mode: Beginning of transient. In this mode, a
trigger pulse is generated at the first step of each sub list.
A sub list may consist of one or more rows as indicated by
the value of the link field. Note that most test scripts
consist of multiple short sub lists to support single step
mode.
List Mode: This mode adds a Trig column to the test
sequence table. A trigger will be generated for each entry
in the trigger column that is set to 1. Any other value
including empty will not generate a trigger out. In this
mode, the user has to set one or more triggers. Setting a
trigger can be done by double clicking on the trigger
column with the left mouse button and entering 1.

Test Status

This area of the Test Control tab displays test execution status for each individual
test step as well as the complete test. The top section shows the complete test
time and what part has been completed so far. The bottom section shows the
same information for the present test step.

Test Description

The area at the top of the Test Control tab is used to display additional
information about the selected test tab. It is for information purposes only and has
no effect on the actual test execution.

Table Header

The table header is shown against a yellow background and shows the table from
which this test sequence was selected.

Test Sequence Data
Grid

The test sequence data grid displays the specific parameters used to program the
power source. This data implements the required tests. The parameters are
loaded for the referenced ini file in the test table on the Test Selection tab. It is
possible to edit this information if needed which allows variations on any of these
tests to be made. Any changes can be saved to a new file. The standard provided
test files are write protected however and should not be overwritten.
The following columns/data fields are available in this table for each test step
(row).

Step number

Sequential number from 1 through end of sequence.

Freq

Test frequency in Hz

FSlew

Frequency slew rate in Hz/sec

Volt A

Test voltage for phase/output A in Vrms or Vdc depending on
mode.