3.1.3.1
Specific Test Limitations and Discrepancies
While every effort has been made to provide comprehensive coverage of the RTCA/DO-160 tests,
some limitations apply, largely due to hardware limitations of the Lx power source such as available
programming resolution or rounding. The following table lists specific implementation issues by test
number. The deviation from the standard requirement is shown in the “Actual setting” column. The
reason for the deviation is shown in the last column.
DC tests are NOT supported by the Lx/Ls Series
Vnom
Test
Std requirement
Actual setting
Reason
115VAC
A(CF):16.5.1.2
115 +/- .18 Vrms
115 +/- .2 Vrms
Resolution
115VAC
A(NF):16.5.1.2
115 +/- .18 Vrms
115 +/- .2 Vrms
Resolution
115VAC
A(WF):16.5.1.2
115 +/- .18 Vrms
115 +/- .2 Vrms
Resolution
115VAC
A(CF):16.5.1.2
115 +/- 1.24 Vrms
115 +/- 1.2 Vrms
Resolution
115VAC
A(NF):16.5.1.2
115 +/- 1.24 Vrms
115 +/- 1.2 Vrms
Resolution
115VAC
A(WF):16.5.1.2
115 +/- 1.24 Vrms
115 +/- 1.2 Vrms
Resolution
230VAC
A(CF):16.5.1.2
230 +/- .36 Vrms
230 +/- .4 Vrms
Resolution
230VAC
A(CF):16.5.1.2
230 +/- 2.48 Vrms
230 +/- 2.5 Vrms
Resolution
115VAC
A(CF):16.5.1.3
400Hz mod by .75Hz 400Hz mod by .8Hz
Resolution
115VAC
A(NF):16.5.1.3
360Hz mod by .75Hz 360Hz mod by .8Hz
Resolution
115VAC
A(NF):16.5.1.3
650Hz mod by .75Hz 650Hz mod by .8Hz
Resolution
115VAC
A(WF):16.5.1.3
360Hz mod by .75Hz 360Hz mod by .8Hz
Resolution
115VAC
A(WF):16.5.1.3
800Hz mod by .75Hz 800Hz mod by .8Hz
Resolution
230VAC
A(CF):16.5.1.3
400Hz mod by .75Hz 400Hz mod by .8Hz
Resolution
230VAC
A(CF):16.5.1.5.1
320V Transient
No deviation but
–HV
option required to run.
Max voltage 320Vrms requires
400V range.
230VAC
A(CF):16.5.2.3.1
360V Transient
No deviation but
–HV
option required to run.
Max voltage 360Vrms requires
400V range.