Lx/Ls Series -704 Option Limitations

MIL-STD-704 rev A through F tests are all implemented via firmware for Lx/Ls Series power sources.
Please see Lx/Ls Series user manual for operation instructions and limitations (P/N 7004-980).

5.2

–704 Power Groups

The

–704 option supports both AC and DC modes of operation. The correct mode is a function of the

EUT to be tested and the operator must select the corresponding operating mode first. This can be
done from the front panel if needed or using the Gui program. Use of the Gui program for all settings
will be assumed for the remainder of this chapter.

Seven power groups are covered in the MIL-STD-704 directive:

Group

Description

SAC

Single Phase 115VAC / 400 Hz (CF)

TAC

Three Phase 115VAC / 400 Hz (CF)

SVF

Single Phase 115VAC / Variable Frequency (VF), 360

– 800 Hz.

TVF

Three Phase 115VAC / Variable Frequency (VF), 360

– 800Hz.

SXF

Single Phase 115VAC / 60 Hz (CF)

LDC

28VDC

HDC

270VDC

The required steady state output settings for each power group must be selected based on the type of EUT to
be tested. These settings can be made from the front panel of the power source or from the main Gui control
screen.

5.3 Test Tables


The tables shown on the next few pages list the test sections covered by the

–704 option. Where

relevant, the numbering used matches that of the MIL-STD-704 document. The notes in the right
hand column indicate tests that are either not directly supported or require additional external test
equipment. Test tables are listed by rev and then by power group. The list starts with the most current
rev to date, rev F.