Avionics Software
6.2
–A350 Power Groups
The
–A350 option supports both AC and DC modes of operation. The correct mode is a function of
the EUT to be tested and the operator must select the corresponding operating mode first. This can
be done from the front panel if needed or using the Gui program. Use of the Gui program for all
settings will be assumed for the remainder of this chapter.
Nine power groups are covered in the ABD0100.1.8.1 A350 directive:
Group
Description
SVF
AC, Single phase, 115VAC, Variable frequency.
SVFH
AC, Single phase, 230VAC, Variable frequency.
SCF
AC, Single phase, 115VAC, Constant frequency.
SCFH
AC, Single phase, 230VAC, Constant frequency.
TVF
AC, Three phase, 115VAC, Variable frequency.
TVFH
AC, Three phase, 230VAC, Variable frequency.
TCF
AC, Three phase, 115VAC, Constant frequency.
TCFH
AC, Three phase, 230VAC, Constant frequency.
LDC
DC 28VDC
The required steady state output settings for each power group must be selected based on the type of
EUT to be tested. These settings can be made from the front panel of the power source or from the
main Gui control screen.
6.3 Test Tables
The tables shown on the next few pages list the test sections covered by the
–A350 option. Where
relevant, the numbering used matches that of the ABD0100.1.8.1 document. The notes in the right
hand column indicate tests that are either not directly supported or require additional external test
equipment. Test tables are listed by rev and then by power group. The list starts with the most current
rev to date, rev B.
Data shown is for MX Series. Other models may have slight variations from the data shown here.