Specific Test Limitations and Discrepancies
While every effort has been made to provide comprehensive coverage of the A350 tests, some
limitations apply, largely due to hardware limitations of the various power source models. The
following table lists specific implementation issues by test number. The deviation from the standard
requirement is shown in the “Actual setting” column. The reason for the deviation is shown in the last
column. Unless indicates otherwise, these discrepancies apply to all models covered by this manual.
Rev Test
Standard requirement
Actual setting
Reason
B
103
0.18 Vrms modulation
0.2 Vrms modulation
Programming resolution
B
103
1.24 Vrms modulation
1.2 Vrms modulation
Programming resolution
B
103H
0.36 Vrms modulation
0.4 Vrms modulation
Programming resolution
B
103H
2.48 Vrms modulation
2.5 Vrms modulation
Programming resolution
B
104
Voltage Spike
Not supported
Requires additional equipment
B
107
Voltage distortion 2 700Hz to 150KHz Not supported
Requires additional equipment
B
CF 110
+/-0.75Hz modulation
+0.7/-0.8 Hz modulation
Programming resolution
B
VF 111
+/-0.75Hz modulation
+0.7/-0.8 Hz modulation
Programming resolution
B
203
0.18 Vrms modulation
0.2 Vrms modulation
Programming resolution
B
203
1.24 Vrms modulation
1.2 Vrms modulation
Programming resolution
B
203H
0.36 Vrms modulation
0.4 Vrms modulation
Programming resolution
B
203H
2.48 Vrms modulation
2.5 Vrms modulation
Programming resolution