beautypg.com

Return loss zero reference measurement – EXFO IQS-12004B DWDM Passive Component Test System for IQS-500 User Manual

Page 96

background image

Calibrating the DWDM Passive Component Test System

84

IQS-12004B

Performing Reference Measurements

Return Loss Zero Reference Measurement

The Return Loss Zero measurement provides a zero reflection reference to
the WRM return loss detector. This reference will be used to compensate
for internal WRM parasitic reflections as well as parasitic reflections
originating at the WRM output/launch fiber connector.

Note: A Return Loss Zero reference measurement is required each time the

system is turned on, and is recommended whenever there is a significant
change in environmental conditions.

To perform the Return Loss Zero reference measurement:
1.
Clean the connectors and connect the modules as indicated on

the screen. For more information, see Maintenance on page 213 and
Connecting Optical Fibers on page 66.

2. Select Return Loss Zero in the IQS-12004B Reference window and

press OK.