COM Objects Reference

278

IQS-12004B

Measures Interface—Methods

Measures

CalWr

Description

This command performs a wavelength response calibration on

the power meter channel.
ChannelToCalibrate represents the number of channels to

calibrate, as indicated on the channel localization list.

Syntax

object.CalWr(ChannelToCalibrate, FireEvent)

Parameters

Byval Long ChannelToCalibrate, Byval Boolean FireEvent

Response

None

Possible error ErrInitBefore, ErrInProgress, ErrChannelNumber, ErrAcqType,

ErrAcqStopped, ErrScanRange, ErrSourceNotEnabled. For more

information, see IQS-12004B DWDM Passive Component Test
System (Standard) COM Error Messages
on page 208.

See also

ChannelLocalisation (see page 291)

Measures

RefIl

Description

This command performs an IL reference measurement.

Syntax

object.RefIl(FireEvent)

Parameters

Byval Boolean FireEvent

Response

None

Possible error ErrInitBefore, ErrCalWrBefore, InProgress, ErrAcqType,

ErrAcqStopped, ErrScanRange, ErrSourceNotEnabled. For more

information, see IQS-12004B DWDM Passive Component Test
System (Standard) COM Error Messages
on page 208.