Technical Specifications

230

IQS-12004B

IQS-12004B MultiPath Testing Option

IQS-12004B MultiPath Testing Option

Specifications

1

Loss measurement uncertainty

2, 3, 5

± 0.06 dB

Loss measurement repeatability

2, 3, 6

± 0.04 dB

PDL uncertainty

2, 4, 5

± (0.05 dB + 5 % of DUT PDL)

PDL repeatability

2, 4, 6

± (0.025 dB + 3 % of DUT PDL)

ORL uncertainty

2, 5, 7

± 1 dB (0 dB to 45 dB)
± 2 dB (45 dB to 50 dB)

ORL repeatability

2, 3, 6

± 0.3 dB (0 dB to 45 dB)

ORL range

2

0 dB to 55 dB

Operating temperature

Any stable temperature at ± 1 °C
in the range 20 °C to 30 °C

All other specifications

Please refer to IQS-12004B data sheet

Standard Accessories

Instruction manual, installation software CD

and interconnecting patchcords

Optional Accessories

Switch module with launch fibers
Rackmount kit
Video monitor

Notes:
1. All specifications are typical, after 60-minute warmup and testing a low

loss components using the recommended EXFO accessories and EXFO
low PDL switch.

2. Using a semi-rigid patchcord between the IQS-9401 and the switch.
3. Valid for spectrally uniform component. Does not include uncertainties

due to connector or connector adapter.

4. Valid for spectrally uniform component terminated with a non-angled

connector. Does not include uncertainties due to connector or
connector adapter.

5. For first 100 cycles, within 100 hours after calibration and reference.
6. For 100 consecutive cycles within 100 hours.
7. Calibrated using a UPC connector with 14.7 dB of return loss.