System overview – EXFO IQS-12004B DWDM Passive Component Test System for IQS-500 User Manual

Page 16

background image

Introducing the IQS-12004B DWDM Passive Component Test System

4

IQS-12004B

System Overview

System Overview

The IQS-12004B DWDM Passive Component Test System provides spectral
insertion loss (IL), polarization-dependent loss (PDL), and optical return
loss (ORL) measurements with high resolution and accuracy.

The DWDM Passive Component Test System sweeps a very-low-noise
tunable laser source across the spectral band of interest (C-band or
C+L-band). While the source is sweeping, synchronized measurements
from the Wavelength Reference Module (WRM) and the measurement
power meters are acquired into the system software. At the end of the
sweep, advanced signal processing is applied to the data, resulting in
accurate IL, PDL, and ORL data.

This configuration ensures quick testing time that is practically
independent of the number of device ports (there is a nominal data
transfer and analysis penalty per channel). Because of the low source
spontaneous emissions (SSE) of the IQS-2600B/IQS-2600CT Tunable Laser
Source, a dynamic range of 60 dB is easily attained. The wavelength
reference module ensures wavelength and loss accuracy by providing
a fast and continuous wavelength and power reference throughout the
sweep.

The numerous software features are explained later in this user guide.

IQS-1643T

IQS-2600B
or
IQS-2600CT
Tunable
Laser
Source

IQS-5150
Polarization
State
Adjuster

IQS-9401
Wavelength Reference
Module

Device
Under Test

Power Reference

ORL

Wavelength
Reference

IQS-1643T