Using the Wavelength Calibration Software

DWDM Passive Component Test System

145

Using Calibration Tools

Using Calibration Tools

There are two ways to access this step:

³

By clicking Tools during the first step.

³

By carrying out the entire process.

³

Apply New: writes the new calibration constants found for the
IQS-9401 WRM module and displays a calibration certificate. For more
information on the calibration certificate, see Viewing and Printing a
Calibration Certificate
on page 146. However
, if the whole process is
not performed, you will not be able to access the Apply New button
because no calibration constants have been found.

³

Apply Old: Accessible even if the whole calibration process is not
performed. It re-writes calibration constants, by applying an old
calibration file or the Factory Default.

³

Save Current: Accessible even if the whole calibration process is not
performed. It allows you to save the current calibration constants to a
binary file identified with the following format name:
SerialNumber-Date-Hour.dat.