Using the Power Meter Linearity Verification Tool

158

IQS-12004B

Setting Up the Acquisition Parameters

Setting Up the Acquisition Parameters

The linearity verification is made by comparing results obtained when
attenuating the power level on a power meter to be tested and a reference
power meter.

The power direction can be decreasing, increasing or both (a specific trace
will be displayed and values will be added to the results grid for each of the
power directions).

If you select both directions, the decreasing values will be displayed first.
The increasing values will be added just after.

Power direction

First power value in

dBm

(for a specific

acquisition)

Last power value in

dBm

(for a specific

acquisition)

Attenuation

Decreasing

Max power value

a

a.

Power value obtained when variable attenuator is set to the lowest (attenuation) value supported

Min power value

b

b.

Minimum power threshold allowed by the application

Increases

Increasing

Min power value

b

Max power value

a

Decreases

Actual power

Minimum power threshold (non-editable)